Outer Box Dimensions: Approximately 1500×1973×1855 (W×H×D) (mm)
Temperature Range:
High-temperature zone: RT ~ 170℃
Low-temperature zone: RT ~ -60℃
Test zone: -40 ~ 150℃
Temperature Shock Range:
High temperature: 60℃ ~ 150℃
Low temperature: -10℃ ~ -40℃
Normal temperature: Indoor temperature
Temperature Shock and Recovery Time:
Temperature shock time: 5S
Temperature recovery time: 5 minutes
Shock process (as shown in the figure)
The green line represents the heat storage curve of the high-temperature zone, the red line represents the shock curve of the test zone, and the blue line represents the cold storage curve of the low-temperature zone.
Temperature Fluctuation: ≤±0.5℃
Temperature Error: ≤±2℃
Temperature Uniformity: ≤±2.0℃ (Temperature uniformity is the arithmetic mean of the difference between the measured maximum and minimum temperatures in each test)
Heating Rate of Preheating Zone: ≥3℃/min (non-linear throughout the process)
Cooling Rate of Precooling Zone: ≥1℃/min (non-linear throughout the process)
Functional Requirements:
Equipment Weight: 550KG
Equipment Power: 380V 32KW
Note: The above performance indicators are measured under the conditions of ambient temperature +25℃, relative humidity ≤85%RH, and no test samples.