Dimensions of Test Area Basket: 500×600×500 W×H×D (mm)
Outer Box Dimensions: Approximately 1200×2263×1650 W×H×D (mm)
Temperature Range:
Heat storage in high-temperature zone: +60℃ ~ +170℃
Cold storage in low-temperature zone: -10℃ ~ -55℃
Temperature Shock Range:
High temperature: 60℃ ~ 150℃
Low temperature: -10℃ ~ -45℃
Temperature Shock and Recovery Time:
Shock time: within 10 seconds
Temperature recovery time of test area: 5 minutes
Shock process (as shown in the figure)
The red line represents the temperature shock curve, the blue line represents the cold storage zone curve, and the green line represents the heat storage zone curve.
Temperature Fluctuation: ≤±0.5℃
Temperature Error: ≤±2℃
Temperature Uniformity: ≤±2.0℃ (Temperature uniformity is the arithmetic mean of the difference between the measured maximum and minimum temperatures in each test)
Heating Rate of Preheating Zone: ≥3℃/min (non-linear throughout the process)
Cooling Rate of Precooling Zone: ≥1℃/min (non-linear throughout the process)
Functional Requirements:
Cooling and heating of indoor space (air);
The indoor space reaches the set temperature according to the temperature setting;
Divided into two parts: high-temperature zone and low-temperature zone;
The two zones are independent, and the temperature is switched through the basket.
Equipment Weight: 600KG
Equipment Power: 380V 28KW
Note: The above performance indicators are measured under the conditions of ambient temperature +25℃, relative humidity ≤85%RH, and no test samples.
This test equipment is prohibited from:
Testing or storing samples of flammable, explosive, or volatile substances;
Testing or storing samples of corrosive substances;
Testing or storing biological samples;
Testing or storing samples of strong electromagnetic emission sources;
Testing or storing radioactive substances;
Testing or storing samples that may produce flammable, explosive, volatile, highly toxic, corrosive, or radioactive substances during testing or storage.